Analogue Bandwidth (50 Ω, -3 dB) | MSO8064: 600MHz MSO8104: 1GHz MSO8204: 2GHz |
Analogue Bandwidth (1 MΩ, -3 dB) | 500 MHz |
Calculated Rising Time under 50 Ω (single-channel mode, 10%-90%, typical) | MSO8064: ≤583 ps MSO8104: ≤350 ps MSO8204: ≤175 ps |
No. of Input/Output Channels | 4 input analogue channels 1 input EXT channel 16 input digital channels (required to purchase the RPL2316 logic analyser probe) dual-channel arbitrary waveform generator output (required to purchase the MSO8000-AWG option) |
Sampling Mode | real-time sampling |
Max. Sample Rate of Analog Channel | 10 GSa/s (single-channel), 5 GSa/s (half-channel), 2.5 GSa/s (all channels) Note: When all the channels are enabled, the sample rate is 2.5 GSa/s, and the analogue bandwidth can reach up to 1 GHz. |
Max. Memory Depth | analogue channel: 500 Mpts (single-channel), 250 Mpts (half-channel), 125 Mpts (all channels) digital channel: 62.5 Mpts (all channels) |
Max. Waveform Capture Rate | ≥600,000 wfms/s |
Hardware real-time waveform recording and playing | ≥450,000 wfms (single-channel) |
Peak Detection | capture 400 ps glitches |
LCD Size and Type | 10.1-inch capacitive multi-touch screen/gesture enabled operation |
Display Resolution | 1024 × 600 |
Vertical System Analogue Channel | |
---|
Input Coupling | DC or AC |
Input Impedance | 1 MΩ ± 1%, 50 Ω ± 1% |
Input Capacitance | 19 pF ± 3 pF |
Probe Attenuation Coefficient | 0.0001X, 0.0002X, 0.0005X, 0.001X, 0.002X, 0.005X, 0.01X, 0.02X, 0.05X, 0.1X, 0.2X, 0.5X, 1X, 2X, 5X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X, 2000X, 5000X, 10000X, 20000X, and 50000X |
Probe Recognition | auto-recognised RIGOL probe |
Maximum Input Voltage | 1 MΩ CAT I 300 Vrms, 400 Vpk, Transient Overvoltage 1600 Vpk 50 Ω 5 Vrms |
Vertical Resolution | 8 bits |
Vertical Sensitivity Range | 1 MΩ 1 mV/div ~ 10 V/div 50 Ω 1 mV/div ~ 1 V/div |
Offset Range | 1 MΩ: ± 1 V ( 1 mV/div ~ 50 mV/div ) ± 30 V ( 51 mV/div ~ 260 mV/div ) ± 100 V ( 265 mV/div ~ 10 V/div ) 50Ω: ±1 V ( 1 mV/div ~ 100 mV/div ) ±4 V ( 102 mV/div ~ 1 V/div ) |
Dynamic Range | ±5 div (8 bits) |
Bandwidth Limit (Typical) | 1 MΩ: 20 MHz, 250 MHz; selectable for each channel 50 Ω: 20MHz |
DC Gain Accuracy | ± 2% of full scale |
DC Offset Accuracy | ≤200 mV/div (±0.1 div±2 mV±1.5% of offset value) >200 mV/div (±0.1 div±2 mV±1.0% of offset value) |
Channel-to-Channel Isolation | ≥ 100:1 (DC to 1 GHz), ≥ 30:1 (> 1 GHz to the rated bandwidth) |
ESD Tolerance | ±8 kV (on input BNCs) |
Vertical System Digital Channel | |
---|
Number of Channels | 16 input channels (D0 ~ D15) (D0 ~ D7, D8 ~ D15) |
Threshold Range | ±20.0 V, in 10 mV step |
Threshold Accuracy | ±(100 mV + 3% of the threshold setting) |
Threshold Selection | TTL(1.4 V), CMOS5.0(2.5 V), CMOS3.3(1.65 V), CMOS2.5(1.25 V), CMOS1.8(0.9 V), ECL(-1.3 V), PECL(3.7 V), LVDS(1.2 V), and 0.0V User (adjustable threshold for 8 channels in a group) |
Max. Input Voltage | ± 40 V peak CAT I; transient overvoltage 800 Vpk |
Max. Input Dynamic Range | ±10 V + threshold |
Minimum Voltage Swing | 500 mVpp |
Input Impedance | about 101 kΩ |
Probe Load | ≈8 pF |
Vertical Resolution | 1 bit |
Horizontal System Analogue Channel | |
---|
Range of Time Base | 600MHz: 500 ps/div~1 ks/div 1GHz: 500 ps/div~1 ks/div 2GHz: 200 ps/div~1 ks/div support fine adjustment |
Time Base Resolution | 2 ps |
Time Base Accuracy | ±1 ppm ± 2 ppm/year |
Time Base Delay Range | before triggering ≥1/2 screen width after triggering 1 s to 100 div |
Time Interval (△T) Measurement | ±(1 sample interval) ± (2 ppm×readout) ± 50 ps |
Inter-channel Offset Correction Range | ±100 ns |
Horizontal Mode | YT: Default XY: X = Channel 1, Y = Channel 2 SCAN: Time base ≥200 ms/div, available to enter or exit the SCAN mode by rotating the Horizontal SCALE knob ROLL: Time base ≥200 ms/div, available to enter or exit the ROLL mode[5] by rotating the Horizontal SCALE knob |
Horizontal System Digital Channel | |
---|
Min. Detectable Pulse Width | 3.2ns |
Maximum Input Frequency | 500 MHz (accurately copied as the sine wave of the maximum frequency of the logic square wave; input amplitude is the minimum swing; the shortest ground cable is required for the logic probe) |
Inter-channel Time Delay | 1 ns (typical), 2 ns (maximum) |
Trigger System | |
---|
Trigger Source | Analogue channel (1 ~ 4), Digital channel (D0 ~ D15), EXT TRIG, and AC Line |
Trigger Mode | Auto, Normal, Single |
Trigger Coupling | DC: DC coupling trigger AC: AC coupling trigger High Frequency Rejection: High frequency rejection, cut-off frequency ~ 75 kHz (internal only) Low Frequency Rejection: Low frequency rejection, cut-off frequency ~ 75 kHz (internal only) |
Noise Rejection | increase delay for the trigger circuit (internal only), On/Off |
Holdoff Range | 8 ns to 10 s |
Trigger Bandwidth | Internal: analogue bandwidth of the oscilloscope External: 200 MHz |
Trigger Sensitivity (Internal) | 1 div, <10 mV/div 0.6 div, 10 mV/div ~ 19.8 mV/div 0.4 div, 20 mV/div ~ 49.5 mV/div 0.35 div, ≥50 mV/div When the noise rejection is enabled, the trigger sensitivity is reduced half |
Trigger Sensitivity (External) | 200 mVpp, DC ~ 100 MHz 500 mVpp, 100 MHz ~ 200 MHz |
Trigger Level Range | Internal: ± 5 div from the center of the screen External: ± 8 V AC Line: fixed 50% |